Language:
    • Available Formats
    •  
    • Availability
    • Priced From ( in USD )
    • Secure PDF 🔒
    • Immediate download
    • $480.00
    • Add to Cart
    • Printed Edition
    • Ships in 1-2 business days
    • $480.00
    • Add to Cart

Customers Who Bought This Also Bought

 

About This Item

 

Full Description

Preface:

This is the first edition of CAN/CSA-C22.2 No. 60601-2-68, Medical electrical equipment — Part 2-68: Particular requirements for the basic safety and essential performance of X-ray-based image-guided radiotherapy equipment for use with electron accelerators, light ion beam therapy equipment and radionuclide beam therapy equipment, which is an adoption, with Canadian deviations, of the identically titled IEC (International Electrotechnical Commission) Standard 60601-2-68 (first edition, 2014-09). It is one in a series of Standards issued by CSA Group under Part II of the Canadian Electrical Code.

For brevity, this Standard will be referred to as "CAN/CSA-C22.2 No. 60601-2-68" throughout.

This Standard is intended to be used in conjunction with CAN/CSA-C22.2 No. 60601-1:14, Medical electrical equipment — Part 1: General requirements for basic safety and essential performance (adopted IEC 60601-1:2005+A1:2012, with Canadian deviations).

201.1.1 Scope:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of X-ray based IMAGE-GUIDED RADIOTHERAPY equipment for use with EXTERNAL BEAM EQUIPMENT (EBE).

This particular standard covers safety aspects of kilovoltage (kV) and megavoltage (MV) X-ray imaging devices in a known geometrical relationship with EBE for the purpose of IGRT. It covers aspects of communication and relationships between the EXTERNAL BEAM EQUIPMENT and X-ray imaging devices, attached or not directly attached to, but in the same RADIATION shielded area as, and dedicated for use only with, the EXTERNAL BEAM EQUIPMENT.

This particular standard deals with equipment for REAL-TIME X-IGRT, ONLINE X-IGRT and OFFLINE X-IGRT. It covers procedures to reduce the risk of over-reliance on the X-IGRT EXTERNAL BEAM SYSTEM (X-IGRT EBS). For example the manufacturer will provide an interactive interface for user interaction with the correction suggested by the system.

If a clause or subclause is specifically intended to be applicable to X-IGRT EBE SYSTEMS the content of that clause or subclause will say so. If that is not the case, the clause or subclause applies only to X-IGRT EQUIPMENT.

This particular standard, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively to the MANUFACTURER and some installation aspects of X-IGRT EBE SYSTEMS intended to be
  • for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED PERSONS by OPERATORS having the required skills for a particular medical application, for particular specified clinical purposes, e.g. STATIONARY RADIOTHERAPY or MOVING BEAM RADIOTHERAPY
  • maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE
  • subject to regular quality assurance performance and calibration checks by a QUALIFIED PERSON.
NOTE In this particular standard, all references to installation refer to installation in the RESPONSIBLE ORGANIZATION?S premises

201.1.2 Object:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCE requirements for X-IGRT EQUIPMENT and X-IGRT EBE SYSTEMS.