1.1 Scope

Replacement:

This Particular Standard specifies requirements for the safety of electrically powered laboratory centrifuges designed to be installed and used in accordance with the Rules of the Canadian Electrical Code, Part I.

1.1.1 Aspects excluded from scope

Addition:

- additional precautions which may need to be observed when centrifuging materials which are flammable or explosive (see 5.4.101); - additional precautions which may need to be observed when centrifuging materials that chemically interact vigorously (see 5.4.101).

1.1.2 Equipment excluded from scope

Addition:

- rotating electrical machinery within the scope of IEC 34; - CENTRIFUGES (used in explosive atmospheres) within the scope of IEC 79.

1.2 Object

Addition:

- contact with moving parts (see 7.2); - CENTRIFUGE movement during any DISRUPTION (see 7.3.101); - expelled parts (see 7.5); - small particles from ROTOR ASSEMBLIES, expelled in air exhaust systems (see 7.5.104); - high energy chemical reaction after ROTOR DISRUPTION (see 7.5.105); - ineffectiveness of BIOSEALS (see 13.101).

1.4 Environmental conditions

Modification:

Modify the third dash as follows:

- temperature +2 degrees C to +40 degrees C.

 

Document History

  1. CAN/CSA C22.2 NO. 61010-2-020:17 (R2022)


    Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-020: Particular requirements for laboratory centrifuges (Adopted CEI/IEC 61010-2-020:2016)

    • Most Recent
  2. CAN/CSA C22.2 NO. 61010-2-020-09 (R2014)


    Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-020: Particular requirements for laboratory centrifuges (Adopted CEI/IEC 61010-2-020:2006, second edition, 2006-05)

    • Historical Version
  3. CAN/CSA C22.2 NO. 1010.2.020-94(R04)

    👀currently
    viewing


    Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 2-020: Particular Requirements for Laboratory Centrifuges